Cases Journal


Open Access Case Report

Failure of asnis iii 5.0 mm cannulated screw: a case report

A Chen*, C Willis-Owen, K Akhtar and S Kamineni

Author Affiliations

Department of Trauma and Orthopaedics, Charing Cross Hospital, Fulham Palace Road, London, UK

For all author emails, please log on.

Cases Journal 2010, 3:9 doi:10.1186/1757-1626-3-9

Published: 7 January 2010

Abstract

We describe a case of an ASNIS III 5.0 mm partially threaded cannulated screw unthreading itself as it was being inserted during fracture fixation of a humerus. The majority of complications associated with cannulated screws involve breaking of the screw, as opposed to unthreading. We believe that the self tapping design of the screw, in combination with the cannulated design, creates a potential area of weakness when used on hard bone.